WaveForms™ SDK Reference Manual
Page 70 of 85
Pulse signal:
For pulse signal the initial level and initial value are specified with FDwfDigitalOutCounterInitSet function. These
are loaded when entering Running state.
10.1 Control
FDwfDigitalOutReset(HDWF hdwf)
Parameters:
- hdwf Interface handle.
The function above resets and configures (by default, having auto configure enabled) all of the instrument
parameters to default values.
FDwfDigitalOutConfigure(HDWF hdwf, BOOL fStart)
Parameters:
- hdwf Interface handle.
- fStart Start the instrument. To stop, set to FALSE.
The function above is used to start or stop the instrument.
Divider
Initial
Divide
Divide
Divide
...
Triggered/Running state
Counter start high
Initial
Low
High
Low
...
Counter start low
Initial
High
Low
High
...
Each expire of divider is one step for the counter